久久综合网久久综合,国产综合久久99久久,五月丁香伊人网,亚洲欧美日韩第一区中文字幕,免费香蕉一区二区在线观看,日韩在线精品视频观看,91资源总站,国产综合91天堂亚洲国产
        客戶咨詢熱線:
        Technical articles技術(shù)文章
        首頁 > 技術(shù)文章 >USP <787> Subvisible Particulate Matter in Therapeutic Protein Injections

        USP <787> Subvisible Particulate Matter in Therapeutic Protein Injections

         更新時間:2015-05-15  點擊量:8232
          

        Follow the same standardization steps as described in USP <788>

        標(biāo)準(zhǔn)制定遵從USP <788>制定步驟

        • Volume accuracy 

        體積
        • Flow rate

        流速
        • Sensor resolution

        傳感器靈敏度
        • Calibration

        校準(zhǔn)
        • Count accuracy

        計數(shù)

        Two system controls tests are recommended

        兩種系統(tǒng)控制測試

        • Blank test: measure five aliquots of degassed particle-free water, the count must be less than 1 particle/mL > 10 µm.

        測試:檢測脫氧純水,> 10 µm的粒子數(shù)不得超過1 個/mL
        • System suitability verification; analyze the USP particle count reference standard and pass the count limits

        系統(tǒng)適應(yīng)性驗證:檢測分析USP標(biāo)準(zhǔn)粒子并低于限定值

        Sample preparation

        樣品準(zhǔn)備

        • If there is enough volume test individual units

        確認每個樣品是否有足夠的體積
        • If the volume is too small, mix units and combine the contents to obtain the required volume (typically 0.2-5.0 mL)

        如果體積太小,將樣品混合直到符合測量的體積(典型0.2-5.0 mL)
        • Degas the sample and gently mix again

        輕輕上下晃動混合均勻

        Analysis

        • Analyze four aliquots

        測量4次
        • Count the particles in the size range of interest, including >10 & 25 µm

        測量計數(shù)規(guī)定范圍內(nèi)的粒子,包括>10 跟 >25 µm。
        • Disregard the first result and average the second two

        *次不計,求得后面兩次的平均值

        Pass/Fail criteria

        If the container volume < 100 mL;  less than 6000 particles/container > 10 µm and 600 particles/container > 25 µm

        通過/不通過的標(biāo)準(zhǔn)范圍

        如果標(biāo)示裝量< 100 mL;> 10 µm 不超過6000個/瓶,> 25 µm 不超過600個/瓶

        If the container volume > 100 mL; less than 25 particles/mL > 10 µm and 3 particles/mL > 25 µm. Also not exceed the per container limits for the < 100 mL  criteria above.

        如果標(biāo)示裝量> 100 mL,>10 µm 不超過25個/ml,> 25 µm 不超過10個/ml

        The AccuSizer FX Nano system is ideal system for performing the USP <787> test for the following reasons

        AccuSizer FX Nano是測量USP <787>的*選擇。

        • Passes all standardization and system suitability tests

        通過所有標(biāo)準(zhǔn)粒子與系統(tǒng)適應(yīng)性測試
        • Sensitivity down to 0.15 µm

        靈敏度可以達到0.15 µm
        • Higher concentration limit

        可檢測高濃度樣品
        • Accurate sample volume down to 250 µL

        小測量體積為250 µL

         

        Aggregated IgG protein before and after filtration as measured on the AccuSizer FX Nano

        AccuSizer FX Nano關(guān)于聚合免疫蛋蛋白經(jīng)過濾和未過濾的檢測譜圖

        滬公網(wǎng)安備 31011202021364號